Laserwelt investiert in neueste Technik!Der neue Tattoolaser "Trivantage" von Syneron/Candela ist...
Kein Aprilscherz!Was gut ist gehört einfach zusammen! Bereits während der AAD (American Academy of...
Wie oft haben Sie sich schon über minderqualifizierte Konkurrenten gewundert, die mit bombastischem...
Laserweltteam has won another University Hospital as new costumer. Prof. Dr. med. Dr. h.c. Thomas...
Frim now on also Prof. Giunta the new Chief of surgery departement of LMU University Munich owns a...
Hans Metelmann, Professor of the University Greifswald and Study Chairman DALM (Diploma in...
Laserwelt has won University Clinic Salzburg ENT as new High End Costumer. Prof. Rasp as Top Chief...
worldwide wellknown Univ.-Prof. Dr. Dr. med. Prof. h.c. (RC) Norbert Pallua and his teamDr. med....
University Hospital Leipzig - Oral and Maxillofacial Surgery is working with Canfield Vectra M3.
as new Canfield Vectra User. Prof. Paasch and Dr. Wagner are very happy to offer the new 3D Vectra...
Laserwelt hat keine Kosten und Mühen gescheut und unterstützt Dr. Klaus Hoffmann auf seiner...
Yes, electrodes may be steam-sterilized at 135° C (for max. 18 min). Important: Please also see the user instructions for ARROWtip™ electrodes.
ARROWtip™-electrodes have a shaft diameter of 2.4 mm. Adaptors for handpieces with a shaft diameter of 4 mm are also available.
Yes, ARROWtip™ electrodes will work with all radiofrequency units as well as with common electrosurgical units.
It is not possible to give a standard reply to this question as each unit has its special characteristics that have to be taken into account. Best results are generally achieved in blend or cut & coag mode with 3 to 8 watts.
For the radiofrequency unit BM-780 II this equals a power setting of 1.5 to 2.5 in Cut II mode.
For cutting, only use the tip of the probe (not the entire needle). Brush the tip of the ARROWtip™-electrode several times across the tissue - similar to when using a brush, but always in the same direction.
Do not use abrasive or sharp-edged objects or abrasive sponges to clean the probes. It is best to use wet swabs to clean the electrodes.
Do not bend ARROWtip™ electrodes.
Protect the tips of the ARROWtip™ electrodes during cleaning and for storing.
Important: Please also see the user instructions for ARROWtip™ electrodes.
If this happens, the power setting chosen is too low. Adjust it gradually upwards until the desired cut comes out clean.
If this happens, the power setting chosen is too high. Adjust it downwards until cutting is possible without charring the tissue.
© Copyright 2011 - Laserwelt ™ - Michael Landsberg. Die Marke Laserwelt ist urheberrechtlich geschützt.
Nicht genehmigtes Kopieren von Texten, Bildern, Dateien ist ausdrücklich untersagt.
info(at)laserwelt.com